PERCUTANEOUS PIGTAIL NEPHROSTOMY SET (2020-08-31) — recall 61595

Health Canada — Medical Devices · Canada · 2020-09-11 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61595
Date of alert
2020-09-11
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. If a catheter is left indwelling longer than 4 weeks, there is an increased potential for breakage, kinking, encrustation, migration, and/or infection to occur. Potential adverse events that may result from these hazards include discomfort, additional intervention, hydronephrosis, and loss of kidney function.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/percutaneous-pigtail-nephrosto

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.