Product Wiki · Recalls · Health Canada — Medical Devices

Heater-Cooler Unit HCU 40 (2020-08-17) — recall 61580

Health Canada — Medical Devices · Canada · 2020-09-04 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61580
Date of alert
2020-09-04
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Maquet Cardiopulmonary GmbH has received complaints in relation to leaking vacuum valves of the HCU 40. Based on engineering testing, it is possible that after performing the emptying mode, the vacuum valve does not fully close. When the HCU 40 is then operated again, water can leak.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/heater-cooler-unit-hcu-40-2020

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.