Product Wiki · Recalls · Health Canada — Medical Devices

VITROS XT 7600 Integrated & 3400 Chemistry System (2020-07-24) — recall 61544

Health Canada — Medical Devices · Canada · 2020-08-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61544
Date of alert
2020-08-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: During internal testing, ortho identified that slot 18 of some cm rotors of the VITROS XT 7600 and XT 3400 systems were not machined adequately, leaving a rough surface that may contribute to the misalignment of a slide at the time of the reflectometer reading. When a microslide is improperly positioned within slot 18 of the cm rotors, a condition code may not post. If an undetected microslide position error occurs, an erroneous test result may be generated.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vitros-xt-7600-integrated-3400

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.