VNS Therapy System — recall 61408

Health Canada — Medical Devices · Canada · 2020-06-26 · Type II

As described by Health Canada — Medical Devices: VNS Therapy System - VNS Therapy Programming Software Model 3000 1.6 (2020-06-10)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61408
Date of alert
2020-06-26
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Reason for recall is response to reports received from the field (first report received on November 11, 2019) that some M103-M106 generators are displaying a false Low Output Current Warning message upon re-interrogation operations using the M3000 software version 1.6 programmer, when the generator is actually providing stimulation as intended. Issue root cause is the software code. French translation can be found in Attachment 1.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vns-therapy-system-vns-therapy

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.