Product Wiki · Recalls · Health Canada — Medical Devices

CLINITUBES with Capillary Caps & Safe CLINITUBES Blood Sampler… — recall 61350

Health Canada — Medical Devices · Canada · 2020-06-12 · Type II

As described by Health Canada — Medical Devices: CLINITUBES with Capillary Caps & Safe CLINITUBES Blood Sampler (2020-03-03)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61350
Date of alert
2020-06-12
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Radiometer Medical ApS has received an increased number of complaints stating that several customers have had troubles when using SafeClinitubes (942-892). The SafeClinitubes have little or no aspiration, which means that it is difficult or impossible to collect the patient blood with the SafeClinitubes. The lack of aspiration may lead to the hazardous situation of delayed availability of fetal scalp pH measurement results during the delivery process.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/clinitubes-capillary-caps-safe

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.