Product Wiki · Recalls · Health Canada — Medical Devices

Alinity C system Optics Assembly (2020-05-20) — recall 61307

Health Canada — Medical Devices · Canada · 2020-05-29 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61307
Date of alert
2020-05-29
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: This Product Correction letter is to inform you of an issue with the Alinity c system Optics Assembly (part number C-35016278-01). Abbott has identified the potential for a nonlinear optics response due to an issue with the Alinity c optics. The issue is with the diffraction grating within the optics assembly and can result in a lower than expected absorbance value.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/alinity-c-system-optics-assemb

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.