Michele Sleep Scoring System (2020-05-04) — recall 61304
Health Canada — Medical Devices · Canada · 2020-05-29 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Vulnerabilities (CVE-2020-11651/CVE-2020-11652) identified in the software deployment tool used by the Michele Sleep Scoring System allowed attackers to access and run arbitrary commands on both the 'master' and linux based 'minion' salt servers on 02- May-20. The attackers used this to stop the MICHELE software on these servers and run cryptocurrency mining software.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.