Benchmark Ultra Automated Immunostaining System & Benchmark XT… — recall 61233

Health Canada — Medical Devices · Canada · 2020-05-08 · Type III

As described by Health Canada — Medical Devices: Benchmark Ultra Automated Immunostaining System & Benchmark XT (2020-04-08)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61233
Date of alert
2020-05-08
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Following the information provided by our manufacturer on investigation of customer complaints, we would like to inform you of an issue that has been identified in Ventana software system (VSS V 11.0 to 12.5.3 when used with identified staining instruments) that can result in the software allowing a dispenser to be used beyond their intended capacity when a slide is aborted/stopped while the SW is in compiling or starting state. The manufacturer advised us of this on 06-Apr-2020.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/benchmark-ultra-automated-immu

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.