Monnal Eva Expiratory Valve, Single use & Expiratory Flow Sensor… — recall 61232
Health Canada — Medical Devices · Canada · 2020-05-08 · Type II
As described by Health Canada — Medical Devices: Monnal Eva Expiratory Valve, Single use & Expiratory Flow Sensor, Single use (2020-03-29)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: These specific measures linked to the current outbreak for the care of Covid-19 patients are based on the growing demand for supply of Monnal Eva UU Expiratory sets and also for the membranes (KY665300) used with the Monnal Eva Expiratory Valve autoclavable version. To ensure continuity of the care pathway during this ramp up of production, and better protect the final user against Covid 19, Air Liquide Medical Systems : authorizes by temporary derogation the disinfection of the Monnal Eva UU Expiratory sets and membranes (5 cycles) recommends to add a hydrophobic filter before the expiratory sets.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.