Product Wiki · Recalls · Health Canada — Medical Devices

VITROS Chemistry Products PHYT Slides (2018-05-10) — recall 61014

Health Canada — Medical Devices · Canada · 2018-06-06 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
61014
Date of alert
2018-06-06
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Ortho confirmed customer complaints regarding the potential for imprecise results when using specific lots of the VITROS PHYT Slides. Ortho performed internal testing to confirm the issue using several lots of coating 0165. The results obtained were either positively or negatively biased across the VITROS PHYT Slides measuring range of 3.00-40.00 ug/ml (11.88-158.40 umol/l).
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/vitros-chemistry-products-phyt

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.