Agency
Health Canada — Medical Devices
Recall identifier
60865
Date of alert
2018-04-24
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Zimmer Biomet is conducting a lot specific recall for the Echo Bi-Metric Collarless Porous Stem (size 7). In 2008, the Echo Bi-Metric Hip Stems underwent a design change to reduce the lateral shoulder of the device in order to make insertion of the device easier during surgery. It was later determined that item 195207 has the potential to fracture based on insertion testing.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/echo-bi-metric-collarless-poro

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.