Infinity M540 (2018-03-19) — recall 60820

Health Canada — Medical Devices · Canada · 2018-04-13 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
60820
Date of alert
2018-04-13
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Draeger learned about an issue with the infinity 540 patient monitor with software versions vg4.1 and below. When the m540 patient monitor is used with and connected to the Infinity MCable Mainstream CO2 (m11.1) sensor revision 16, the m540 patient monitor will reboot if the sensor enters a reduced accuracy mode. If the m540 patient monitor reboots 3 times within a 10 minute period or less, the m540 patient monitor will go enter a fail-state, resulting in a resetting to factory defaults and loss of patient settings and stored patient data. Revision 16 sensors were introduced in mid-2017. Use of revision 15 and earlier sensors will not cause this issue to occur.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/infinity-m540-2018-03-19

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.