Graphic Cases for 6.5 mm and 7.3 mm Cannulated Screw Sets (2018-03-06) — recall 60774
Health Canada — Medical Devices · Canada · 2018-03-23 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: DePuy Synthes is conducting a lot-specific recall of the graphic cases for 6.5mm and 7.3mm cannulated screw sets because they contain screw gauges that incorrectly measure screws 5.41mm shorter than the actual overall screw length. Thus, when using the measuring gauge, the surgeon would potentially be inserting a screw which is 5.41mm shorter than intended.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.