Product Wiki · Recalls · Health Canada — Medical Devices

Giraffe Incubator, Giraffe Omnibed, Giraffe Incubator Carestation… — recall 60072

Health Canada — Medical Devices · Canada · 2019-11-01 · Type II

As described by Health Canada — Medical Devices: Giraffe Incubator, Giraffe Omnibed, Giraffe Incubator Carestation, and Giraffe Omnibed Carestation (2019-10-22)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
60072
Date of alert
2019-11-01
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The bedside panels of Giraffe Incubator, Giraffe Omnibed, Giraffe Incubator Carestation, and Giraffe Omnibed Carestation can be upright and look closed but not be latched. The portholes also can look closed when not latched. If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. If a bedside panel or porthole that is not latched falls open, it will no longer protect the patient from falling.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/giraffe-incubator-giraffe-omni

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.