SynchroMed II - Programmable Pump (2019-10-11) — recall 60045
Health Canada — Medical Devices · Canada · 2019-10-22 · Type I
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Medtronic is voluntarily retrieving specific Synchromed II Implantable Drug Infusion Pumps after investigating complaints related to permanent motor stall. This voluntary recall is being conducted due to the potential for the presence of a foreign particle inside the pump motor assembly which could interfere with motor gear rotation and lead to a permanent motor stall. The source of the foreign particle has been identified and eliminated. A permanent pump motor stall will result in cessation of drug infusion therapy which may result in return of underlying symptoms and/or withdrawal symptoms.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.