Product Wiki · Recalls · Health Canada — Medical Devices

BACT/ALERT VIRTUO - A & B UNIT (2019-10-21) — recall 59936

Health Canada — Medical Devices · Canada · 2019-11-01 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59936
Date of alert
2019-11-01
Country / jurisdiction
Canada
Authority classification
Type II
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: A problem was reported by a customer where, on a BACT/ALERT VIRTUO R2.0 with MYLA, patient culture bottles of various types were "discharged at 7 days instead of 10 days". The customer reported that the maximum test time (MTT) of each bottle was modified at the LIS from seven (7) days to ten (10) days; however, the instrument only tested the bottles for seven days and then discarded them to the waste bin. The Biomerieux investigation could not duplicate the described behavior. However, evaluation concluded the root cause to be a VIRTUO software/firmware timing anomaly whereby the processing of multiple software actions, following a new bottle scan, within a one (1) second time interval may lead to a MTT discrepancy.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/bactalert-virtuo-b-unit-2019-1

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.