Health Canada — Medical Devices · Canada · 2019-08-02 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: It was observed that some Magna Pure 24 Sample Racks do not properly hold the sample tubes in place during sample processing. When a sample tube is not properly positioned (i.e. slanted) within the rack, it could lead to an instrument crash, pipette tips missing the sample tubes during sample aspiration, and possible incorrect sample volume measurements caused by early liquid level detection (ELLD) events.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.