MAGNA PURE 24 System (2019-07-08) — recall 59843

Health Canada — Medical Devices · Canada · 2019-08-02 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59843
Date of alert
2019-08-02
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: It was observed that some Magna Pure 24 Sample Racks do not properly hold the sample tubes in place during sample processing. When a sample tube is not properly positioned (i.e. slanted) within the rack, it could lead to an instrument crash, pipette tips missing the sample tubes during sample aspiration, and possible incorrect sample volume measurements caused by early liquid level detection (ELLD) events.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/magna-pure-24-system-2019-07-0

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.