ADVIA Centaur System - Folate Assay (2019-07-29) — recall 59819
Health Canada — Medical Devices · Canada · 2019-08-16 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Siemens Healthcare Diagnostics has confirmed the potential for the ADVIA Centaur homocysteine assay (100 test kit smn 10310374; 500 test kit smn 10310375) to intermittently interfere with the folate assay when testing is performed on the ADVIA Centaur, ADVIA Centaur XP or ADVIA Centaur XPT systems. The interference is observed intermittently with only the first replicate of folate when the assay is tested immediately following homocysteine assay testing.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.