Health Canada — Medical Devices · Canada · 2019-08-16 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Biohorizons determined certain lots of the AutoTac Handle of 400-200 instrument were assembled with a non-conforming reduced size spring spacer. The non-conforming spring spacer of .049 Inches was assembled in the handle instead of the conforming .230 Inches spring spacer.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.