V60 Ventilator, Symbols Only, with CFLEX and AVAPS Options… — recall 59673
Health Canada — Medical Devices · Canada · 2019-10-03 · Type II
As described by Health Canada — Medical Devices: V60 Ventilator, Symbols Only, with CFLEX and AVAPS Options (2019-09-18)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The touchscreen must be used to change certain therapy settings. The touchscreens in affected units may become frozen and fail to respond to touch commands. Consequently, if a patient requires a change in a ventilator therapy setting, e.g., FIO2, the change could not be made. Although it would be immediately apparent to the user that the changes cannot be made, there is no advance warning of touchscreen failure.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.