Product Wiki · Recalls · Health Canada — Medical Devices

Spot Vision Screener (2018-06-18) — recall 59665

Health Canada — Medical Devices · Canada · 2019-10-11 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59665
Date of alert
2019-10-11
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: If the device is not used or charged every five months, the coin cell battery that maintains the system clock may become depleted. This may cause the patient age to be calculated incorrectly. An incorrect patient age could result in an incorrect (either under or over) referral recommendation.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/spot-vision-screener-2018-06-1

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.