GentleCool Pro Cryogen Canister (2019-09-09) — recall 59648
Health Canada — Medical Devices · Canada · 2019-10-11 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: The issue observed in the field and documented in Candela's associated complaints details that the neck bushing (or top) of the canister becoming disconnected from the body of the canister, causing rapid depressurization. This event can injure the patient/operator by a) uncontrolled cryogen gas spray and b) propelling the gas canister away from the system in an uncontrolled manner with significant velocity. Please refer to the attached health hazard evaluation for a complete description.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.