Agency
Health Canada — Medical Devices
Recall identifier
59648
Date of alert
2019-10-11
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The issue observed in the field and documented in Candela's associated complaints details that the neck bushing (or top) of the canister becoming disconnected from the body of the canister, causing rapid depressurization. This event can injure the patient/operator by a) uncontrolled cryogen gas spray and b) propelling the gas canister away from the system in an uncontrolled manner with significant velocity. Please refer to the attached health hazard evaluation for a complete description.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/gentlecool-pro-cryogen-caniste

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.