MultiDiagnost-ELEVA, OmniDiagnost Systems (2019-03-03) — recall 59526

Health Canada — Medical Devices · Canada · 2019-03-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59526
Date of alert
2019-03-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: For the height and tilt movement of the stand/table, the MultiDiagnost-ELEVA, OroDiagnost and OmniDiagnost systems have 2 actuators that are connected to the system stand by 4 shafts. The position of each of these shafts is secured by a locking plate. During service of two MultiDiagnost-ELEVA systems, it was noticed that a locking plate was missing. In one of these cases, the shaft was still properly positioned after the system had been in use for 9 years. In the other case, involving a system that had been in use for 5 years, the shaft was no longer properly positioned but was still holding the actuator. Over time, a missing locking plate may result in the shaft of the corresponding connection point becoming improperly positioned. If the shaft works its way totally out of position, it could result in unexpected stand/table tilt movement that cannot be stopped by the user.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/multidiagnost-eleva-omnidiagno

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.