Health Canada — Medical Devices · Canada · 2019-04-12 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: During post market surveillance, it was realized that some customers are using flammable freezing sprays within the Cryostat chamber. Such flammable freezing sprays can potentially ignite and can cause serious injuries. The current device labeling does not include any specific warning about the usage of flammable freezing sprays, therefore the corrective action will improve the device labeling with an adhesive warning label and safety amendment to the instrument instruction for use (IFU).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.