WS80A Diagnostic Ultrasound System, HS70A Ultrasonic Diagnostic… — recall 59358

Health Canada — Medical Devices · Canada · 2019-07-05 · Type II

As described by Health Canada — Medical Devices: WS80A Diagnostic Ultrasound System, HS70A Ultrasonic Diagnostic Apparatus (2019-06-24)

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59358
Date of alert
2019-07-05
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: A specific mode of operation of certain versions of WS80A/HS70A systems may pose a risk to patient safety. On June 13, 2019, the manufacturer discovered that probes may overheat when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode on WS80A/HS70A systems. There has been no actual injury reported to Samsung Medison.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/ws80a-diagnostic-ultrasound-sy

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.