Product Wiki · Recalls · Health Canada — Medical Devices

Stratus CS Acute Care System Cardiac troponin-i TestPak (2019-06-04) — recall 59220

Health Canada — Medical Devices · Canada · 2019-06-21 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59220
Date of alert
2019-06-21
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Increased rate of false positive cTnI results beginning 2 months prior to the end of the product's shelf life due to a positive bias and increased imprecision. Increased rate of random non-repeatable false positive cTnI results at any time in the product's shelf life.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/stratus-cs-acute-care-system-c

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.