Product Wiki · Recalls · Health Canada — Medical Devices

microTargeting Drive System, STar Drive System (2018-04-29) — recall 59191

Health Canada — Medical Devices · Canada · 2019-05-31 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
59191
Date of alert
2019-05-31
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: microTargeting Depth Stop is being recalled which is a component of microTargeting STar Drive System and microTargeting Drive System. It has been sold individually and as part of the system. After analysis of several complaints of lead damage it was determined that, the user overtightened the depth stop screw onto the lead, despite the warning in DFU. FHC determined to replace depth stops proactively and determined field corrective action is the best way.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/microtargeting-drive-system-st

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.