Health Canada — Medical Devices · Canada · 2019-05-10 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: Device operating on software v1.7.0 may display intermittent contact force data during procedures. The device log data fills the allocated disk space, which prevents the storage on new log data. In such instances, the contact force information is displayed in purple indicating that the data is not current and may not be accurate. This data will remain purple until communication is re-established or the disk has room to store log files.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.