Sysmex Urine Quality Control and Calibrator (2018-11-19) — recall 58885
Health Canada — Medical Devices · Canada · 2018-12-10 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: Sysmex received a complaint from their customers regarding failing Quality Control (QC) for the Red Blood Cells (RBC), White Blood Cells (WBC), Epithelial Cells (EC), casts (CAST) and bacterial (BACT) with the UF II Control. The control failures were for one or a combination of parameters that recovered values on the high side of the printed assay ranges. The manufacturer of the Sysmex UF-1000i has confirmed that the UF II Control and UF II Calibrator Assay targets and limits for the lot numbers YS8029, YS8032, YS8033, and S8023 have been incorrectly assigned during the manufacturing process. This incorrect assignment may cause QC results to recover outside the printed assay ranges.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.