Product Wiki · Recalls · Health Canada — Medical Devices

HeartStart MRX Rechargeable Lithium Ion Battery (2018-10-27) — recall 58882

Health Canada — Medical Devices · Canada · 2018-11-30 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58882
Date of alert
2018-11-30
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Certain MRx M3538A lithium-ion batteries may contain a defective component. Should the component fail, the battery will no longer charge or deliver power to the MRx Monitor/Defibrillator, which may prevent operation of the device if a second charged battery is not installed or if the device is not connected to AC or DC power. Additionally, the LED status indicator on the battery will not illuminate, even if the battery has a charge. Should a second, unaffected battery be present in the HeartStart MRx, the battery failure may go unnoticed.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/heartstart-mrx-rechargeable-li

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.