Sapphire Epidural Set Yellow-Striped Microbore (2018-11-04) — recall 58877
Health Canada — Medical Devices · Canada · 2018-11-28 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: ICU Medical has identified an issue with the yellow striped sapphire epidural set. It was identified that the inclusion of both a 0.2 micron filter and a Pressure Activated Valve (PAV) with the filter in a proximal location can increase the probability of distal occlusions alarms. This configuration is limited to this sapphire epidural set.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.