Reported risk: Philips Volcano is initiating a voluntary field corrective action to address an interoperability issue that affects Volcano s5i, core and core mobile systems with version 2.5 FFR/iFR software connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS). Other Hemodynamic Systems which do not adhere to commonly used pressure calibration values (0 v @ 0mmhg and 1v @ 100 mmhg) may also be impacted. The version 2.5 FFR/iFR software is designed to calibrate with the aortic output of the Hemodynamic Monitoring System (HMS). If the HMS adheres to commonly used pressure calibration values, then the interoperability issue will not have any impact. However, if the system does not adhere to commonly used values the interoperability issue will prevent calibration of the software with an impacted HMS. In cases where the HMS does not adhere to commonly used values, the volcano pa (ao) and the HMS ao may be misaligned by up to +/- 5 mm hg. This discrepancy may lead to confusion. Philips has confirmed that the FFR/iFR normalization process ensures there is negligible impact to the FFR/iFR measurements.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.