Health Canada — Medical Devices · Canada · 2018-11-28 · Type II
As described by Health Canada — Medical Devices: Brilliance 40/60 Slice CT System, 6/10/16 Slice CT System, and CT System (16P) (2018-10-23)
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: Philips has identified an issue with the front gantry cover from a particular supplier, which may impact the front gantry cover of your CT system. Some covers from this supplier may develop a problem where the fiberglass weakens at the connection point to the gantry frame causing the front cover to separate from the gantry frame, allowing it to fall.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.