Health Canada — Medical Devices · Canada · 2018-11-28 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: A deviation was found in measuring performance for above listed test strip products, when measured on Urisys 1100 Urine Analyzer. The performance for Limit of Detection (LoD) of parameter Protein, Nitrite, Leukocytes and Erythrocytes does not correspond to the values listed in the respective registration documents. The discrepant LoD can lead to false negative results at the limit of detection level, for affected parameters (Protein, Nitrite, Leukocytes or Blood (intact erythrocytes)).
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.