Chemstrip 5, Chemstrip 7 and Urisys 1100 (2018-08-27) — recall 58856

Health Canada — Medical Devices · Canada · 2018-11-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58856
Date of alert
2018-11-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: A deviation was found in measuring performance for above listed test strip products, when measured on Urisys 1100 Urine Analyzer. The performance for Limit of Detection (LoD) of parameter Protein, Nitrite, Leukocytes and Erythrocytes does not correspond to the values listed in the respective registration documents. The discrepant LoD can lead to false negative results at the limit of detection level, for affected parameters (Protein, Nitrite, Leukocytes or Blood (intact erythrocytes)).
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/chemstrip-5-chemstrip-7-and-ur

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.