Product Wiki · Recalls · Health Canada — Medical Devices

Captus 3000 and 700 (2018-11-04) — recall 58834

Health Canada — Medical Devices · Canada · 2018-11-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58834
Date of alert
2018-11-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Capintec received reports of arm failures on Captus 3000 systems shipped in 2011-2012. This malfunction will cause the arm and collimator assembly (45 pounds) to fall to its lowest position, which has the potential for injury if the collimator comes in contact with operator or patient.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/captus-3000-and-700-2018-11-04

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.