Product Wiki · Recalls · Health Canada — Medical Devices

Creatine kinase-MB (CK-MB) (2018-08-05) — recall 58753

Health Canada — Medical Devices · Canada · 2019-01-11 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58753
Date of alert
2019-01-11
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Following internal investigation of customer complaints, the manufacturer confirmed that an increased absolute absorbance for both calibration standards and reduced control recoveries could occur for the product Creatine Kinase-MB (#05168562190) lot # 31195601 used on cobas c 701/702 module.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/creatine-kinase-mb-ck-mb-2018-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.