Tremoflo C-100 Airwave Oscillometry System (2018-09-10) — recall 58506
Health Canada — Medical Devices · Canada · 2018-10-19 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: On given limited number of Tremoflo C-100 Airwave Oscillometry Systems, there may be a mismatch between the hardware configuration and the device calibration factor that can cause a reversal of the flow signal, which may also lead to a reversal of some of the intra-breath parameters calculated by the software.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.