Product Wiki · Recalls · Health Canada — Medical Devices

Stockert Heater Cooler System 3T (2018-10-28) — recall 58482

Health Canada — Medical Devices · Canada · 2018-11-15 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58482
Date of alert
2018-11-15
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: LivaNova is providing updated instructions to monitor the concentration of hydrogen peroxide in the water circuit of the 1T and 3T Heater-Cooler devices, to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm. This regimen enhances the water maintenance regimen outlined in Section 6.4 of the operating instructions.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/stockert-heater-cooler-system-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.