O-arm O2 Imaging System (2018-10-14) — recall 58472
Health Canada — Medical Devices · Canada · 2018-11-15 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: Medtronic has developed a software update (SW version 4.1.0) that will address potential software issues/anomalies that resulted in unused X-Ray dose, delays in therapy (in all cases, of less than 1 hour), patient exposure to non-navigated surgery, and navigational inaccuracy. Medtronic has also received complaints associated with Mobile View Station (MVS) fuses within the ac power input circuit. This is resolved with the new design of the AC power input circuit. Medtronic will install the component(s) for the new AC power input circuit and IAS product label application and update the user manual.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.