Product Wiki · Recalls · Health Canada — Medical Devices

Entroy Pool Lift (2018-03-16) — recall 58387

Health Canada — Medical Devices · Canada · 2018-09-28 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58387
Date of alert
2018-09-28
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: A potential for the chair or stretcher detachment from Entroy pool lifter is caused by a fact that caregiver may suppose that the pin has been paired with the seat/stretcher, but there could be a misalignment that might be unnoticed by the caregiver so the seat/stretcher does not get fully paired. Instead, once the lift reaches a certain height, the seat/stretcher falls off the lift. To raise awareness and reduce risk a quick reference guide will be installed with the Entroy pool lift device.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/entroy-pool-lift-2018-03-16

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.