Product Wiki · Recalls · Health Canada — Medical Devices

DX-D 600 (2018-08-15) — recall 58371

Health Canada — Medical Devices · Canada · 2018-08-31 · Type II

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58371
Date of alert
2018-08-31
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: AGFA has been notified by the manufacturer of their DX-D 600 device that one incident occurred on a similar dr device. The incident was caused by the breakage of the two steel cables that support the equipment to its roof anchor. To their knowledge this was the first occurrence of such an incident. Due to this, part of the equipment might fall down and can potentially harm the patient, user or third parties. This problem could occur in the most unfavorable case of excessive wear of the vertical pulley in the column.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/dx-600-2018-08-15

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.