Product Wiki · Recalls · Health Canada — Medical Devices

CyPass Micro-Stent Model 241 (2018-08-30) — recall 58370

Health Canada — Medical Devices · Canada · 2018-09-07 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58370
Date of alert
2018-09-07
Country / jurisdiction
Canada
Risk level
Type I
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Alcon is conducting a Voluntary Field Safety Corrective Action (withdrawal) of all versions of the CyPass Micro-Stent. The COMPASS-XT study was designed to collect safety data on the subjects who participated in the COMPASS study for an additional three years, with analysis of the completed data set at five years post-surgery. At five years, the CyPass Micro-Stent group experienced statistically significant endothelial cell loss compared to the group who underwent cataract surgery alone.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/cypass-micro-stent-model-241-2

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.