Dimension Sample Probe Cleaner (2018-09-27) — recall 58275

Health Canada — Medical Devices · Canada · 2018-10-19 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58275
Date of alert
2018-10-19
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: Siemens is notifying customers of the observation of the error message "sample probe cleaner not detected" (error code 569) during system check and leaking of sample probe cleaner bottle caps. Siemens has confirmed a low frequency (less than 1%) of defective bottles resulting in this error.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/dimension-sample-probe-cleaner

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.