RAPIDPoint 405, 500 and RAPIDLab 1245 Blood Gas Analyzer (2018-07-24) — recall 58187
Health Canada — Medical Devices · Canada · 2018-08-17 · Type II
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Reported risk: Siemens Healthcare Diagnostics has determined that 1 mg/ml and 2 mg/ml of hydroxocobalamin may interfere with Total hemoglobin (tHb) and some of the co-ox fractions that are reported on the RAPIDPoint 405, RAPIDPoint 500, RAPIDLab1245 and RAPIDLab1265 blood gas analyzers. There have not been any reports of injury due to this issue. Siemens is issuing an Urgent Field Safety Notice (UFSN) to inform customers of this potential interfering substance.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.