DSX- Revelation software 6.08-6.28 (2018-08-03) — recall 58144

Health Canada — Medical Devices · Canada · 2018-09-17 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58144
Date of alert
2018-09-17
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The DSX analyzer utilizes operating software Revelatlon. This software use set up as default with an algorithm for dealing with over/ under conversion. If an end user turns this off and they use the over limit in their Quality Control Equation the software will place 0.00 in the equation leading to an incorrect QC interpretation and potential incorrect value assigned to the result.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/dsx-revelation-software-608-62

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.