Agency
Health Canada — Medical Devices
Recall identifier
58104
Date of alert
2018-07-19
Country / jurisdiction
Canada
Risk level
Type II
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: The following issue is being addressed by GE healthcare file no. FMI 26859: GE healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC (volts alternating current) to service engineers working inside the gantry. This issue does not affect patients or operators external to the equipment.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/optima-ct660-and-revolution-ev

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.