Product Wiki · Recalls · Health Canada — Medical Devices

5008 Cordiax HD-PAED option (2018-06-26) — recall 58084

Health Canada — Medical Devices · Canada · 2018-07-13 · Type I

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked on this page. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58084
Date of alert
2018-07-13
Country / jurisdiction
Canada
Authority classification
Type I
Notice type
recall
Product category
Medical devices
Product type
Medical devices
Reported risk: In the unlikely event of a leakage in the hydraulic system, balancing errors of up to 125 ml/h are possible. For patients < 17 kg body weight this could potentially lead to circulatory dysregulation with potential overhydration or underhydration and subsequent impact on cardiovascular stability.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/5008-cordiax-hd-paed-option-20

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.