AQUIOS CL Flow Cytometer System (2018-01-29) — recall 58064
Health Canada — Medical Devices · Canada · 2018-03-02 · Type I
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Reported risk: If the following sequence of events is followed, it will result in sample misidentification and erroneous results will be generated. The system will not identify or flag the erroneous results. The stat button is selected within AQUIOS CL Software and the user opens the single-tube loader door. User creates a test request when prompted, places a specimen tube in single tube loader, and closes the single-tube loader door. the stat button is selected again and the user opens the single-tube loader door within 15 seconds or less of loading the first sample and before the "details" column on running screen changes to "prepping" for that sample. User removes the first specimen tube and replaces with a different specimen tube. If another request is entered for the new specimen and the single-tube loader door is closed, the new specimen tube will be prepared twice. Both test requests will be processed with 2nd specimen tube. If another request is not entered for new specimen and the single-tube loader door is closed, the new specimen tube will be prepared once. The first test request will be processed with 2nd specimen tube.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.