Product Wiki · Recalls · Health Canada — Medical Devices

BiPAP A40 Ventilatory Support System (2017-11-03) — recall 58032

Health Canada — Medical Devices · Canada · 2017-11-30 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
58032
Date of alert
2017-11-30
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Philips Respironics has identified an issue where a non-conforming part (side panel) was utilized on the BiPAP A40 that affects the mechanical connection between the BiPAP A40 and the detachable battery module. The issue results in the inability to connect the BiPAP A40 device to the detachable battery module during setup before putting the ventilator into operation on a user.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/bipap-a40-ventilatory-support-

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.