SOMATOM Definition Systems (2017-11-07) — recall 57992
Health Canada — Medical Devices · Canada · 2017-11-24 · Type III
🏛Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare became aware of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans on P.A. (posterior-anterior) or A.P. topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate. Potentially, the CARE Dose4D software will select the maximum tube current for the uppermost part of the skull, thus leading to unnecessary radiation exposure.
Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.
ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.