SOMATOM Definition Systems (2017-11-07) — recall 57992

Health Canada — Medical Devices · Canada · 2017-11-24 · Type III

Compiled from an official source — the facts on this page are reproduced from the issuing authority’s own published notice, which is linked above. No AI-generated text appears on this page. How we use AI →
Agency
Health Canada — Medical Devices
Recall identifier
57992
Date of alert
2017-11-24
Country / jurisdiction
Canada
Risk level
Type III
Risk type
recall
Product category
Medical devices
Product type
Medical devices
Notice status
Archived by the issuing authority — kept for reference
Reported risk: Siemens Healthcare became aware of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans on P.A. (posterior-anterior) or A.P. topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate. Potentially, the CARE Dose4D software will select the maximum tube current for the uppermost part of the skull, thus leading to unnecessary radiation exposure.
Primary source: https://recalls-rappels.canada.ca/en/alert-recall/somatom-definition-systems-201

Verify each fact at the issuing agency before acting. Recall scope, affected lots, and remediation instructions may be updated by the agency after our last sync.

ProductGuru indexes recall and safety-alert records from ~30 public agency registries. We are not lawyers, regulators or product-safety consultants. This page restates what the named agency published; consult the primary source above (or qualified counsel in your jurisdiction) before any commercial, medical or safety decision.